Late Surgical Conversion After Thoracic Endograft Failure due to Fracture of the Longitudinal Support Wire

Dittmar Böckler, Hendrik von Tengg‐Kobligk, H. Schumacher, S. Ockert, Matthias Schwarzbach, Jens‐Rainer Allenberg

Journal of Endovascular Therapy · 2005 · 29 citations · 14 references

Concepts

Abstract

PURPOSE: To report complications from a thoracic endograft wire fracture and early experience with elective conversion after thoracic endografting. CASE REPORT: A 43-year-old man underwent urgent endovascular repair of a symptomatic post-traumatic thoracic aneurysm in 1999. The patient had been involved in a car accident 14 years before. He developed clinical and radiological signs of graft infection 46 months after stent-graft implantation. Multidetector computed tomography confirmed a fracture of the longitudinal support wire in the Excluder thoracic stent-graft. Additionally, radiological signs of suspected endograft infection were described. Due to concerns over a potential chronic infection, the stent-graft was successfully excised, and a polyester graft was implanted 50 months after primary endovascular repair. CONCLUSIONS: Recognition or strong suspicion of endograft infection requires conversion with removal of the device. Long-term follow-up after endografting is necessary to assess material fatigue that undermines the durability of these implants.