2003 · 154 citations · 24 references
EngineeringBiomedical EngineeringElectrical EngineeringLow-power DesignMicroelectronicsNeural EngineeringAnalog Integrated CircuitsBiomedical InstrumentationCmos Circuit DesignLow-noise Circuit DesignMicroelectrode ArraysSignal AmplificationInput-referred NoiseLow-noise AmplificationLow-noise AmplifiersLow-power Amplifier DesignLow-power Cmos DesignNeural RecordingNeural Signal AmplificationVlsi Implementation
There is a need among scientists and clinicians for low-noise, low-power biosignal amplifiers capable of amplifying signals in the mHz to kHz range while rejecting large dc offsets generated at the electrode-tissue interface. The advent of fully-implantable multielectrode arrays has created the need for fully-integrated micropower amplifiers. We designed and tested a novel bioamplifier that uses a MOS-bipolar pseudo-resistor to amplify signals down to the mHz range while rejecting large dc offsets. We derive the theoretical noise-power tradeoff limit - the noise efficiency factor - for this amplifier and demonstrate that our VLSI implementation approaches that limit. The resulting amplifier, built in a standard 1.5/spl mu/m CMOS process, passes signals from 0.1mHz to 7.2kHz with an input-referred noise of 2.2/spl mu/Vrms and a power dissipation of 80/spl mu/W while consuming 0.16mm/sup 2/ of chip area.
24
Christian Enz, Fran�ois Krummenacher, Eric A. Vittoz · Analog Integrated Circuits and Signal Processing · 1995 · 1.2K citations · Full text
Engineering, Electrical Engineering, Electronic Materials +13