Journal of Prosthodontics · 2006 · 106 citations · 17 references
DentistryImplant DentistryProsthodonticsImplant ProsthodonticsProstheticsProsthodontic MaterialsCasting ProceduresProsthetic MaterialsAbutment Screw LooseningAbutment Screw TorqueImplant Abutment MaterialsImplant-abutment InterfaceScrew Loosening AssessmentScrew Loosening PreventionTitanium Abutments
BACKGROUND: Screw loosening of implant restorations continues to be a complication in implant prosthodontics. Screw joints are subjected to a loss of initially applied torque because of friction and component misfit. It has been suggested that the loss of applied torque is less in machined metal abutments than in cast plastic abutments. PURPOSE: This study compared the loss of applied torque (detorque) values in machined titanium and in cast UCLA-type abutments for external hex abutment/implant interface. MATERIALS AND METHODS: Four groups of 12 samples each were evaluated: (1) machined titanium abutments, (2) premachined palladium abutments cast with palladium, (3) plastic abutments cast with nickel-chromium, and (4) plastic abutments cast with cobalt-chromium. Each abutment was torqued to 30 Ncm according to the manufacturer's instructions and detorqued three times. The mean loss of applied torque (detorque) was recorded as a percentage of the torque applied. Group means were calculated and compared using ANOVA and Tukey's LSD test. RESULTS: Mean detorque values were (1) 92.3 +/- 2.9%, (2) 81.6 +/- 5.0%, (3) 86.4 +/- 4.6%, and (4) 84.0 +/- 7.0%. Machined abutments demonstrated significantly greater detorque values compared with all cast groups (p < 0.05). No significant differences were found among cast groups. CONCLUSION: Machined abutments retained a significantly greater percentage of torque compared with cast abutments. Casting procedures decrease the percentage of applied torque, which may influence final screw joint stability.
17
Forces and moments on Branemark implants.
Bo Rangert, Torsten Jemt, L Jörnéus · PubMed · 1989 · 624 citations
Elongation and preload stress in dental implant abutment screws.
James E. Haack, Ronald L. Sakaguchi, Ting Sun et al. · PubMed · 1995 · 319 citations
Dentistry, Implant Dentistry, Mechanical Behavior Of Materials +12
The effect of implant/abutment hexagonal misfit on screw joint stability.