IEEE Transactions on Nuclear Science · 2001 · 118 citations · 23 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImagingMedical ImagingNuclear Medicine ImagingPlanar ScintigraphyRadiographic ImagingCoded Aperture ImagingDetector DesignGamma Camera ImagingGamma Camera SystemsIn Vivo StudiesSignal-to-noise RatioSignal-to-noise Ratio Enhancement
As compared to pinholes and collimators, coded aperture cameras present the potential advantage of increased signal-to-noise ratio (SNR). This advantage can be used to improve the resolution of conventional imagers while still producing clear images. In this paper, we describe a near-field coded aperture camera with a field of view of 9 /spl times/ 9 cm and 1.7-mm system resolution based on the use of an existing gamma-camera. Experimental results include planar in vivo studies of /sup 99m/Tc-labeled compounds in a mouse.
23
New family of binary arrays for coded aperture imaging
Stephen R. Gottesman, E. E. Fenimore · Applied Optics · 1989 · 532 citations
Coded aperture imaging in X- and gamma-ray astronomy
E. Caroli, J. B. Stephen, G. Di Cocco et al. · Space Science Reviews · 1987 · 400 citations
Natural Sciences, Astronomy, Physics +14
Uniformly redundant arrays: digital reconstruction methods
E. E. Fenimore, T. M. Cannon · Applied Optics · 1981 · 149 citations