A dual optic accommodating foldable intraocular lens

Stephen D. McLeod

British Journal of Ophthalmology · 2003 · 102 citations · 16 references

DOIFull text

Open access

Concepts

Abstract

AIM: To design an accommodating intraocular lens with extended accommodative range that can be adapted to current standard phacoemulsification and endocapsular implantation technique. METHOD: Ray tracing analysis and lens design; cadaver eye implantation. RESULTS: Ray tracing analysis indicated that axial movement of an exaggerated converging anterior optic linked by spring loaded haptics to a compensatory static diverging posterior optic produced greater change in conjugation power of the eye compared to axial movement of a single optic lens. A dual optic one piece foldable silicone lens was constructed and implanted via a 4 mm corneal incision into the capsular bag of two cadaver eyes. CONCLUSION: A dual optic intraocular lens design can increase the optical effect of a given displacement and suggests improvements for accommodating intraocular lenses.

References