Publication | Closed Access
Monopoles for microwave catheter ablation of heart tissue
17
Citations
7
References
2002
Year
Unknown Venue
EngineeringBiomedical EngineeringNormal SalineElectromagnetic CompatibilityComputational ElectromagneticsCardiologyRadiologyCatheter AblationAntenna TestingAntennaMicrowave AntennaMicrowave MeasurementMicrowave EngineeringRadiofrequency HeatingAntenna DesignMetal-tip MonopoleElectrophysiologyFinite-element MethodMultiband Antennas
The finite-element method is applied to the study of monopole antennas for microwave catheter ablation. Three geometries are considered: open-tip, dielectric-tip and metal-tip. Calculations are made for the magnetic field, the reflection coefficient and the power deposition pattern of the antennas immersed in normal saline. The theoretical results agree with measurements performed on prototypes. The differences between the antennas suggest that the metal-tip monopole best fulfills the requirements of catheter ablation of the endocardium. Optimal dimensions for a metal-tip monopole are also presented.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
| Year | Citations | |
|---|---|---|
Page 1