Device for occlusion of rectovaginal fistula: clinical trials.

B H Lee, Du Hwan Choe, Jae Hoon Lee, K H Kim, Do Yeon Hwang, So Yeon Park, C K Cho, S Y Chin

Radiology · 1997 · 15 citations · 0 references

Concepts

Abstract

PURPOSE: To evaluate a device designed for occlusion of rectovaginal fistula in patients with acquired rectovaginal fistula. MATERIALS AND METHODS: The device consisted of a disk portion, an anchoring portion, and a shaft that connected each portion. The device was framed with a nitinol wire. The disk portion was framed in a four-leaf clover configuration, and the anchoring portion had two ellipsoid arms. The disk and anchoring portions were mounted with a nylon patch and coated with silicone. The device was designed for transrectal insertion and was placed in seven women with rectovaginal fistula caused by pelvic irradiation or pelvic surgery. RESULTS: All devices were placed successfully, and all fistulas were completely occluded after placement of the device. No procedural complications were encountered. The follow-up period was 1-26 months. One patient died of distant metastasis 5 months after placement of the device. None of the patients reported leakage from the rectovaginal fistula during the follow-up period. CONCLUSION: The occlusion device may be promising and beneficial for use in occlusion of rectovaginal fistula that results from pelvic irradiation or pelvic surgery.