Clinical Chemistry · 1983 · 42 citations · 11 references
Natural SciencesBiochemistryEndocrinologyAnalytical ChemistryMass Spectrometric AnalysisSteroid BiochemistryBioanalytical Method ValidationEndocrine AssessmentEndocrine TestingHormonal EvaluationInstrumental AnalysisMass SpectrometryEndocrine AssaysGas Chromatography-mass SpectrometryIsotope DilutionIsotope Dilution AnalysisIsotope Dilution Mass SpectrometryMass Spectrometric QuantificationSerum Cortisol MeasurementStable Isotope Dilution Analysis
We describe procedures for measuring cortisol in plasma and serum by isotope dilution and mass spectrometry. A method that incorporated solvent extraction, derivatization, and gas chromatography/high-resolution mass spectrometry provided data of good precision; interassay CVs were generally 3 to 4% for the concentration range of 100-650 nmol/L. Replacing solvent extraction with extraction on a column of Lipidex 1000 or extraction by immunoadsorption yielded data in excellent agreement with the first method. Plasma and serum pools were analyzed to provide target data for use in the U.K. National External Quality Assessment Scheme for cortisol assays. Routine laboratory assays, as judged by comparison with mass-spectrometric data, were generally positively biased except for analysis of a charcoal-stripped plasma supplemented with cortisol. The results emphasize the importance of using unadulterated plasma or serum pools in assessments of steroid assay procedures.
11
A simple radioimmunoassay for plasma cortisol
J. Seth, Lyndesay M. Brown · Clinica Chimica Acta · 1978 · 98 citations
Plasma cortisol determination by mass fragmentography
Ingemar Björkhem, Rolf Blomstrand, Olle Lantto et al. · Clinica Chimica Acta · 1974 · 52 citations