Plastic Microchip Electrophoresis for Analysis of PCR Products of Hepatitis C Virus

Yu‐Hung Chen, Wei-Chang Wang, Kung-Chia Young, Ting‐Tsung Chang, Shu‐Hui Chen

Clinical Chemistry · 1999 · 69 citations · 22 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: Electrophoresis on polymeric rather than glass microstructures is a promising separation method for analytical chemistry. Assays on such devices need to be explored to allow assessment of their utility for the clinical laboratory. METHODS: We compared capillary and plastic microchip electrophoresis for clinical post-PCR analysis of hepatitis C virus (HCV). For capillary electrophoresis (CE), we used a separation medium composed of 10 g/L hydroxypropyl methyl cellulose in Tris-borate-EDTA buffer and 10 micromol/L intercalating dye. For microchip electrophoresis, the HCV assay established on the fused silica tubing was transferred to the untreated polymethylmethacrylate microchip with minimum modifications. RESULTS: CE resolved the 145-bp amplicon of HCV in 15 min. The confidence interval of the migration time was <3.2%. The same HCV amplicon was resolved by microchip electrophoresis in <1.5 min with the confidence interval of the migration time <1.3%. CONCLUSION: The polymer microchip, with advantages that include fast processing time, simple operation, and disposable use, holds great potential for clinical analysis.