Radiographics · 2013 · 59 citations · 54 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography Dose ManagementComputed Tomography PhysicsGastrointestinal ImagingMedical ImagingPatient Dose OptimizationData Acquisition ParametersDose Reduction TechniquesImage Acquisition ParametersIterative ReconstructionOptimization Of Scan ParametersRadiation Dose ManagementRadiation Dose ReductionScan Length Optimization
Computed tomographic (CT) enterography is a diagnostic examination that is increasingly being used to evaluate disorders of the small bowel. An undesirable consequence of CT, however, is patient exposure to ionizing radiation. This is of particular concern with CT enterography because patients tend to be young and require numerous follow-up examinations. There are multiple strategies to reduce radiation dose at CT enterography, including adjusting acquisition parameters, reducing scan length, and reducing tube voltage or tube current. The drawback to dose reduction strategies is degradation of image quality due to increased image noise. However, image noise can be reduced with commercial iterative reconstruction and denoising techniques. With a combination of low-dose techniques and noise-control strategies, one can markedly reduce radiation dose at CT enterography while maintaining diagnostic accuracy.
54
THE TRANSFORMATION OF POISSON, BINOMIAL AND NEGATIVE-BINOMIAL DATA
F. J. Anscombe · Biometrika · 1948 · 1.3K citations
Radiation dose reduction in computed tomography: techniques and future perspective
Lifeng Yu, Xin Liu, Shuai Leng et al. · Imaging in Medicine · 2009 · 395 citations
Medicine, Pediatrics, Radiology +17