<sup>14</sup>C AMS at Suerc: Improving QA Data with the 5MV Tandem and 250KV SSAMS

Philip Naysmith, Gordon Cook, Stewart P.H.T. Freeman, E. M. Scott, Robert F. Anderson, Sheng Xu, Elaine Dunbar, Graham K.P. Muir, A. Dougans, Klaus M. Wilcken,

Radiocarbon · 2010 · 33 citations · 5 references

DOIFull text

Open access

Concepts

Abstract

In 2003, a National Electrostatics Corporation (NEC) 5MV tandem accelerator mass spectrometer was installed at SUERC, providing the radiocarbon laboratory with 14 C measurements to 4–5‰ repeatability. In 2007, a 250kV single-stage accelerator mass spectrometer (SSAMS) was added to provide additional 14 C capability and is now the preferred system for 14 C analysis. Changes to the technology and to our operations are evident in our copious quality assurance data: typically, we now use the 134-position MC-SNICS source, which is filled to capacity. Measurement of standards shows that spectrometer running without the complication of on-line δ 13 C evaluation is a good operational compromise. Currently, 3‰ 14 C/ 13 C measurements are routinely achieved for samples up to nearly 3 half-lives old by consistent sample preparation and an automated data acquisition algorithm with sample random access for measurement repeats. Background and known-age standard data are presented for the period 2003–2008 for the 5MV system and 2007–2008 for the SSAMS, to demonstrate the improvements in data quality.