IEEE Transactions on Medical Imaging · 1991 · 100 citations · 5 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic RadiologyNeuroradiologyBrain ImagingMedical ImagingNeuroimaging PhysicsPositron Emission TomographyPositron Emission Tomography PhysicsCount Rate AnalysisCount Rate PerformanceGainMaximum GainPet QuantificationPositron Emission Tomography InstrumentationSolid PhantomsThree-dimensional Positron Emission Tomography
The noise-equivalent count-rate (NEC) performance of a neuro-positron emission tomography (PET) scanner has been determined with and without interplane septa on uniform cylindrical phantoms of differing radii and in human studies to assess the optimum count rate conditions that realize the maximum gain. In the brain, the effective gain in NEC performance for three-dimensions (3-D) ranges from >5 at low count rates to approximately 3.3 at 200 kcps (equivalent to 37 kcps in 2-D). The gains of the 3-D method assessed by this analysis are significant, and are shown to be highly dependent on count rate and object dimensions.
5
Three dimensional reconstruction of PET data from a multi-ring camera
David W. Townsend, T. Sprinks, Terry Jones et al. · IEEE Transactions on Nuclear Science · 1989 · 124 citations