Medical Physics · 1997 · 63 citations · 32 references
Changes in magnetic resonance (MR) signals during interstitial microwave heating are reported, and correlated with simultaneously acquired temperature readings from three fiber-optic probes implanted in a polyacrylamide gel phantom. The heating by a MR-compatible microwave antenna did not interfere with simultaneous MR image data acquisition. MR phase-difference images were obtained using a fast two-dimensional-gradient echo sequence. From these images the temperature-sensitive resonant frequency of the 1H nuclei was found to decrease approximately by 0.008 ppm/ degree C. The method and results presented here demonstrate that noninvasive MR-temperature imaging can be performed simultaneously with interstitial microwave thermal treatment.
32
Superconducting open-configuration MR imaging system for image-guided therapy.
John F. Schenck, F.A. Jolesz, Peter Roemer et al. · Radiology · 1995 · 437 citations
Interactive MR-guided biopsy in an open-configuration MR imaging system.
Stuart G. Silverman, B. Collick, Michael R. Figueira et al. · Radiology · 1995 · 328 citations