A method for registration of single photon emission computed tomography (SPECT) and computed tomography (CT) images for liver stereotactic radiotherapy (SRT)

20

Citations

10

References

2012

Year

Abstract

It is feasible to use radiolabeled and radiopaque dual body surface markers for registration of SPECT and CT images acquired in separate sessions allowing conformal avoidance of SPECT-defined functional normal liver. Point-based rigid registration accuracy on the patient surface of 6.1 mm can be achieved using 4 dual body surface markers. The main contribution to the registration error is the deformation of the abdominal surface, arising from the inability to setup the patient in the exact same position at different times on two different imaging systems, and to properly account for breathing artifacts on the CT and SPECT images.