Publication | Closed Access
Syringe pumps take too long to give occlusion alarm
23
Citations
6
References
2005
Year
Our results showed that activation of occlusion alarms on both pumps takes a considerable time, and that the mean time to alarm activation was longer with low infusion rates and larger syringes. To reduce occlusion alarm delays smaller sized syringes with low compliance should be used and staff be alerted when using low flow rates with highly concentrated potent drugs.
| Year | Citations | |
|---|---|---|
Page 1