An Approach for the Use of Doppler Ultrasound as a Screening Tool for Hemodynamically Significant Stenosis (Despite Heterogeneity of Doppler Performance)
93
Citations
29
References
1996
Year
The performance of Doppler ultrasound was highly variable. This suggests that Doppler performance is likely overstated in the literature, but specific devices may perform satisfactorily to detect individuals with hemodynamically significant stenosis. Because performance differs substantially among devices, local investigators are strongly urged to maintain local standardization series. With such standardization, ultrasound performance is sufficient for admission to clinical trials and as the is sufficient for admission to clinical trials and as the basis for carotid surgery. However, without quality control many ultrasound machines are not adequate to accurately predict the degree of carotid stenosis and should not be the only test to decide whether surgery is warranted.