The Brewer reference triad

Vitali Fioletov, J. B. Kerr, C. T. McElroy, D. I. Wardle, Vladimir Savastiouk, T. S. Grajnar

Geophysical Research Letters · 2005 · 134 citations · 1 references

Concepts

Abstract

It has been more than 20 years since the Brewer reference triad was established by Environment Canada at Toronto. The triad serves as a reference for traveling standard instruments that are used to calibrate Brewer spectrophotometers around the world. The members of the triad are calibrated on a regular basis at Mauna Loa, Hawaii. Regular tests made with an internal quartz halogen lamp make it possible to track the instrument response between the calibrations. A new analysis of available column ozone data records indicates that the uncertainty in the daily values derived from each instrument is approximately 0.6%. The random errors of individual observations are within ±1% for 90% of all measurements. Sources of potential errors in the individual Brewer measurements as well as quality control tools are also discussed.