Applied Optics · 2002 · 11 citations · 31 references
EngineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingAdaptive OpticsBreast ImagingBreast UltrasonographyTissue OpticsAbsorption In Heterogeneous MediaInstrument CalibrationInterference LocalizationLocalization Accuracy EstimationObject Localization
The amplitude cancellation method that uses dual out-of-phase sources (a phased array system) can sensitively detect and locate small objects in turbid media. The balance of these two sources is crucial to the system's detection sensitivity and accuracy. We describe a convenient method with which to adaptively calibrate the amplitudes of the two sources at each scanning position by use of low-frequency modulation of the intensity of the in-phase and the antiphase sources. We achieve accurate localization ability of the phased array system by accounting for the influence of asymmetrical boundaries and the heterogeneous background absorption. Experimental data on human breast phantoms demonstrate that localization accuracy within several millimeters has been accomplished through this method.
31
Optical tomography in medical imaging
Simon Arridge · Inverse Problems · 1999 · 2.5K citations
Spectroscopy and Imaging with Diffusing Light
Arjun G. Yodh, Britton Chance · Physics Today · 1995 · 801 citations
Engineering, Biophysics, Physics +17