Corrosion at the Cone/Taper Interface Leads to Failure of Large-diameter Metal-on-metal Total Hip Arthroplasties

H. Meyer, Tina Mueller, Gesine Goldau, Kathrin Chamaon, Marcel Ruetschi, Christoph H. Lohmann

Clinical Orthopaedics and Related Research · 2012 · 154 citations · 40 references

DOIFull text

Open access

Abstract

These findings suggest that in modular cone/taper connections, friction of the MoM articulations may cause failure of the cone/taper interface leading to galvanic corrosion and loosening. It is unclear whether the design of this MoM system provides sufficient stability at the taper.

References

40