Journal of Analytical Atomic Spectrometry · 1996 · 49 citations · 17 references
Natural SciencesChemistryToxicologyAnalytical ChemistryAnalytical ToxicologyMass Spectrometric AnalysisEnvironmental Analytical ChemistryHeavy Metal AnalysisHeavy Metal ContaminationInstrumental AnalysisMass SpectrometryMercury ToxicologyCoastal Water QualityDetection Limit AssessmentHeavy Metal DetectionInstrument Detection LimitMethod Detection LimitPreconcentrationReference MaterialsSample Preconcentration
Trace concentrations of mercury in water samples were determined by a method involving a preconcentration procedure and inductively coupled plasma mass spectrometry (ICP-MS) detection. Mercury vapour, generated by tin(II) chloride as the reductant, was trapped by amalgamation on a gold–platinum gauze, released by controlled heating and detected by ICP-MS. A flow injection sample introduction system with time-based injection was used and the sensitivity was found to be proportional to the mass of mercury introduced. Quantitative recoveries of mercury from open ocean sea-water, coastal sea-water and fresh water reference materials were obtained. A detection limit of 200 pg l–1 for a 25 ml sample was obtained. The precision of the method was 1%(relative standard deviation) at 1 µg l–1.
17
On-line separation and preconcentration in flow injection analysis
Zhaohai Zhu, Shukun Xu, Lei Guo et al. · Analytica Chimica Acta · 1988 · 81 citations