Applied Physics Letters · 2011 · 123 citations · 12 references
MedicineBioengineeringBiomedical EngineeringBiomedical ImagingUltrasonographyUltrasoundBlood-brain Barrier DeliveryContrast-enhanced UltrasoundTransducer TechnologyUltrasonicsUltrasound ImagingUltrasound PhysicsBlood-brain Barrier PermeabilityBlood-brain Barrier Permeability ImagingCavitation DetectionCavitation DynamicsCerebral UltrasoundFocus Ultrasound-mediated Bbb DisruptionTranscranial Ultrasound ImagingUltrasound Contrast Agents
In vivo transcranial and noninvasive cavitation detection with blood-brain barrier (BBB) opening in nonhuman primates is hereby reported. The BBB in monkeys was opened transcranically using focused ultrasound (FUS) in conjunction with microbubbles. A passive cavitation detector, confocal with the FUS transducer, was used to identify and monitor the bubble behavior. During sonication, the cavitation spectrum, which was found to be region-, pressure-, and bubble-dependent, provided real-time feedback regarding the opening occurrence and its properties. These findings demonstrate feasibility of transcranial, cavitation-guided BBB opening using FUS and microbubbles in noninvasive human applications.
12
Noninvasive MR Imaging–guided Focal Opening of the Blood-Brain Barrier in Rabbits
Kullervo Hynynen, Nathan McDannold, Natalia Vykhodtseva et al. · Radiology · 2001 · 1.5K citations