Publication | Closed Access
Temperature elevation by HIFU in<i>ex vivo</i>porcine muscle: MRI measurement and simulation study
45
Citations
29
References
2014
Year
The present study demonstrated that for peak temperatures below 85-90 °C numerical simulation results are in excellent agreement with the experimental data in three dimensions. Both temperature rise and lesion size can be well predicted. Due to nonlinear effect the temperature in the focal region can be increased compared with the linear case. The current magnetic resonance imaging (MRI) resolution is not sufficient. Due to the inevitable averaging the measured temperature can be 10-30 °C lower than the peak temperature. Computational fluid dynamics can provide additional important information that is lost using a state of the art MRI device.