Radiology · 1998 · 22 citations · 0 references
Veterinary MedicineRadiologyVeterinary Diagnostic ImagingDiagnostic ImagingDiagnostic RadiologyThoracic ImagingChest RadiographyComputed TomographySmall Animal Thoracic ImagingThoracic RadiographyBronchoscopyImage Acquisition QualityRadiographic Image QualitySection Thickness Optimization
The authors evaluated the relative importance of the following scanning parameters at computed tomographic bronchoscopy in an anesthetized adult sheep's thorax: section thickness (2, 4, 8 mm), pitch (1.0, 1.5, 2.0), milliampere setting (100, 175, 250 mA), and overlap of reconstructed sections (0%, 25%, 50%, 75%). Five blinded readers ranked the images twice in comparison with photographs of the mounted specimen. Differences in image quality were significant (P < .001) with section thickness of 2 mm and a pitch of 1.0. The milliampere setting had only a minor effect on image quality, and a 50% overlap of reconstructed sections was best.