A digital radiology imaging system: description and clinical evaluation

AW Templeton, SJ Dwyer, GG Cox, KR Lee, JA Johnson, N. Martin, CH Chang, WH Anderson, KS Hensley, J. Bialek

American Journal of Roentgenology · 1987 · 18 citations · 12 references

Concepts

Abstract

We have successfully interfaced an advanced microcomputer system to a 47-cm image intensifier for direct digital radiology. Our development effort has included the design and construction of hardware and the implementation of software. The system matrix (spatial resolution) is 1024 X 1024 and contrast resolution is 8 bits deep. A receiver-operator-characteristic analysis comparing conventional and digital laser chest radiographs for the detection of lung nodules demonstrated no statistical difference. Dosimetry measurements identified decreased patient dose. The intensifier-based digital imaging system is being increasingly used for a variety of examinations including IV pyelograms, all kinds of barium studies, and skeletal, trauma, and chest studies. Hard-copy laser-printed digital images with "bones black" are preferred by the radiology staff and have been well accepted by referring clinicians.