EngineeringMaterials EngineeringMaterials ScienceDiagnostic ImagingRadiation DetectionThin FilmsGamma-ray DetectionRadiographic ImagingSemiconductor DetectorsX-ray DetectionX-ray DetectorsX-ray OpticsDark CurrentElectrical PropertiesElectrical Properties AnalysisPhysical Vapor Deposition MaterialsX-ray Detector ArraysX-ray DiagnosticsX-ray Imaging PhysicsX-ray Imaging Systems
We report x-ray imaging results on polycrystalline HgI<SUB>2</SUB> detector used for direct x-ray imaging. Due to its good electrical properties and high stopping power for x-rays and gamma rays, the material is a good candidate for many applications in medical imaging. The deposition of the HgI<SUB>2</SUB> thick films is made by hot wall physical vapor deposition, (PVD) method and some of the structural features are described here. The x-ray response and some dark current data measured on some recently prepared detectors are reported. Some results obtained with poly-HgI<SUB>2</SUB> thin film deposited on an amorphous-Si TFT's imaging array 4' X 4' performed at the Ginzton Technology Center is reported for the first time. Also phantom images received with a similar deposited poly-HgI<SUB>2</SUB> thin film deposited on an amorphous- Si TFT's imaging array 2' X 2' performed at Xerox-PARC Research Center is also given here. The status of HgI<SUB>2</SUB> technology will be discussed.