Optics Letters · 2004 · 44 citations · 10 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsBiophotonicsFiber OpticsOptical CoherenceOptical Coherence TomographyOptical FibersOptical ImagingOptical TomographyFiber Optic InterferometryMichelson InterferometerNumerical ApertureOptical Coherence Tomography InstrumentationPower DistributionReal-time Imaging
High-resolution optical coherence tomography demands a large detector bandwidth and a high numerical aperture for real-time imaging, which is difficult to achieve over a large imaging depth. To resolve these conflicting requirements we propose a novel multifocus fiber-based optical coherence tomography system with a micromachined array tip. We demonstrate the fabrication of a prototype four-channel tip that maintains a 9-14-microm spot diameter with more than 500 microm of imaging depth. Images of a resolution target and a human tooth were obtained with this tip by use of a four-channel cascaded Michelson fiber-optic interferometer, scanned simultaneously at 8 kHz with geometric power distribution across the four channels.
10
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
In vivo ultrahigh-resolution optical coherence tomography
Wolfgang Drexler, Uwe Morgner, Franz X. Kärtner et al. · Optics Letters · 1999 · 931 citations