Design of Differential X-Ray Filters for Low-Intensity Scattering Experiments

Jack A. Soules, W. L. Gordon, C. H. Shaw

Review of Scientific Instruments · 1956 · 30 citations · 6 references

Concepts

Abstract

The conditions for the thicknesses of the foils of a Ross differential x-ray filter are examined to yield the best accuracy of monochromatic intensity measurement for a given heterogeneous beam of x-rays. The results are applied to the problem of monochromatizing the Kα radiation from a copper target x-ray tube using nickel and cobalt foils. The improvement in the differential filter as a monochromator when used with a proportional counter and single-channel pulse-height discriminator is pointed out.

References

6