Rapid Communications in Mass Spectrometry · 1992 · 29 citations · 1 references
Natural SciencesChemistryAnalytical ChemistryMass Spectrometric AnalysisSpectroscopic AnalysisInstrumental AnalysisMass SpectrometryMass Spectrometry Data AnalysisPeptide AnalysisPeptide IdentificationHigh-resolution Mass AnalysisMass Spectral Data AnalysisMass Spectral DeconvolutionMass Spectral ProcessingMass Spectrometric Identification
Abstract The MaxEnt technique has premviously been successfully applied to the deconvolution of electrossppray mass spectra. The latest version of the Cambridge University software, ‘MemSys5’, has now been applied to high resolution mass spectra. Initial results have shown that peaks requiring an instrument resolution of almost 200 000 to separate them are readily resolved by MaxEnt on data acquired at a resolution of only 50, 000. Moreover, the MaxEnt results are accompanied by quantitative and realistic error bars.
1