Analytical Letters · 1980 · 55 citations · 13 references
EngineeringBiochemistryBioengineeringBiosensorsElectrochemistryEnzymologyElectrochemical BiosensorsElectrochemical DetectionElectrochemical SensorsEnzymatic BiosensorsEnzyme KineticsEnzyme-based BiosensorsAmperometric DetectionAmperometric Enzyme BiosensorsGlucose BiosensingGlucose BiosensorsGlucose OxidaseGlucose SensingGlucose SensorsPeroxidase Enzymes
Abstract A glucose measuring device based on the oxidation of glucose by glucose oxidase and an amperometric kinetic detection was developed. The characteristics obtained with this instrument are comparable with the present glucose instruments but the stability of the enzyme membrane is better and the measuring frequency is higher. In order to expand the applicability of this device to other substrates there was developed a family of bioenzyme electrodes. Enzymes producing glucose as enzymes consuming glucose in addition to glucose oxidase were used. For determination of peroxidase substrates besides a peroxidase-catalase electrode a three-enzyme system consisting of glucose oxidase, peroxidase and catalase was used. Key Words: Determination of Saccharides and CofactorsRienzyme ElectrodesGlucose Measuring Device
13
Stuart J. Updike, G. P. Hicks · Nature · 1967 · 1.3K citations
Construction and study of electrodes using crosslinked enzymes
G. Broun · Analytical Chemistry · 1975 · 107 citations
Enzyme electrode sensing oxygen for uric acid in serum and urine
M. Nanjo, G. G. Guilbault · Analytical Chemistry · 1974 · 106 citations