Prototypes of National Ignition Facility neutron time-of-flight detectors tested on OMEGA

V. Yu. Glebov, C. Stöeckl, T. C. Sangster, S. Roberts, G. J. Schmid, R. A. Lerche, M. J. Moran

Review of Scientific Instruments · 2004 · 83 citations · 9 references

Concepts

Abstract

Prototypes of several National Ignition Facility (NIF) neutron time-of-flight (nTOF) detectors have been built and tested on OMEGA. One group uses a plastic scintillator coupled with a microchannel plate (MCP) photomultiplier tube (PMT), either a single-stage (gain up to 103) MCP PMT and a two-stage (gain up to 106) MCP PMT. Two ultrafast scintillators—BC-422 and BC-422Q—were used. Another nTOF prototype is based on a synthetic diamond wafer produced by the chemical vapor deposition. The nTOF detectors were tested on DD (2.45 MeV) and DT (14.1 MeV) neutron-producing implosions on OMEGA. Based on the results of these tests, a set of nTOF detectors is proposed for use on the NIF to measure ion temperature and DD and DT neutron yields from 109 to 1019.

References

9