Magnetic Resonance in Medicine · 2000 · 14 citations · 28 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingDiffusion ImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMagnetic Resonance Signal FormationDiffusion CoefficientsDiffusion MeasurementsMolecular DiffusionScan Time ReductionSpin EchoSpin Echo ImagingSpin Echo SequencesT2 Relaxation
A single spin‑echo experiment can acquire many images of the same slice with different T2 and diffusion weightings, allowing simultaneous measurement of diffusion coefficients and T2 or more precise diffusion values after T2 correction. The technique produces diffusion values in rat brain that agree with conventional methods, reduces scan time, increases the number of diffusion weightings, decreases motion and susceptibility artifacts, and auto‑co‑registers T2 and diffusion images, but has lower SNR, measures only in‑plane diffusion, and is limited to single‑slice acquisition. Published in Magn Reson Med 44:737–745 (2000) and © 2000 Wiley‑Liss, Inc.
A quantitative imaging sequence has been developed to exploit the intrinsic sensitivity of Burst NMR data to molecular diffusion. In the scan time of a single spin echo experiment, it is possible to acquire many images of the same slice, with a different T2 and diffusion weighting. Under favorable conditions, it is possible to obtain both the diffusion coefficient and T2 from the same experiment; or, by correcting for T2 relaxation using a control image, more precise diffusion coefficients may be measured. The quantitative values in rat brain are in agreement with those from conventional experiments. The major gains of this method are the potentially reduced scan time, the higher number of acquired images corresponding to different diffusion weightings, the reduced sensitivity to inter-scan motion artifact and to local variations in magnetic susceptibility, and an automatic co-registration between T2 and diffusion images. Problems with the sequence include a lower signal-to-noise ratio than is achievable with diffusion-weighted spin-echo imaging, the limitation of measuring only in-plane components of diffusion and, at present, single-slice acquisition. Magn Reson Med 44:737–745, 2000. © 2000 Wiley-Liss, Inc.
28
Fast magnetic resonance diffusion‐weighted imaging of acute human stroke
Steven Warach, Daisy Chien, W. Li et al. · Neurology · 1992 · 881 citations