Concepedia

Publication | Closed Access

Anatomy of a defective barrier

58

Citations

0

References

1992

Year

Abstract

Significantly high glove leak rates were noted after surgical and dental procedures, indicating that the present day latex surgical gloves can become an incompetent barrier once they are used. Unused latex surgical gloves demonstrated a higher rate of defects than allowed by the Food and Drug Administration standards, indicating substantial noncompliance of quality control standards by manufacturers as well as inadequate governmental oversight. Double gloving, or the use of thicker latex surgical gloves, would probably reduce the frequency of glove leaks. Latex surgical gloves should be tested for patency before use and during surgical and dental procedures.