British Journal of Applied Physics · 1957 · 23 citations · 7 references
Natural SciencesMaterials SciencePhysicsElectronic MaterialsPlasma PhysicsThermal AnalysisElectron MicroscopyLight MicroscopyEvaporation RatesFilament MaterialsMeasuring LivesOperating ConditionsSolid AngleThermal EvaporationThermionic Emission
The lives of the tungsten filaments used in electron microscopes can be limited by the thermal evaporation of the metal, or by chemical attack, in inferior vacua. The lives to be expected when either cause is dominant are calculated for a variety of operating conditions. Comparison of these calculated values with measured lives shows that thermal evaporation limits lives in microscopes used at high magnifications. The life limited by evaporation decreases as the temperature and thermionic emission increase, and so corresponding values of filament life and ideal gun output brightness (thermionic emission density per unit solid angle) have been calculated for normal working temperatures, in order that a compromise may be made.
7