Noninvasive glucose monitoring in vivo with an optical heterodyne polarimeter

Chien Chou, Chien-Yuan Han, Wen‐Chuan Kuo, Yeu-Chuen Huang, Ching-Mei Feng, Jenn-Chyang Shyu

Applied Optics · 1998 · 90 citations · 12 references

Concepts

Abstract

An amplitude-sensitive optical heterodyne polarimeter was set up to monitor noninvasively the aqueous glucose concentration in a rabbit's eye. A Zeeman laser in conjunction with a Glan-Thompson analyzer was used to generate an optical heterodyne signal. The amplitude of the heterodyne signal linearly related to the optical rotation angle of the aqueous glucose. The concentration of the aqueous glucose in a rabbit's eyeball was measured in vivo. There was a 30-min time delay between observations of aqueous glucose and blood glucose. The detection capability and the reproducibility of the experiment are demonstrated and discussed.

References