Review of Scientific Instruments · 1971 · 18 citations · 1 references
EngineeringAerospace EngineeringAerodynamicsInstrumentationSensorsAerodynamic TestingFlow SensorsTurbulenceTurbulent FlowVibration MeasurementWing AerodynamicsCalibration ProceduresSignal-to-noise RatioSignal-to-noise Ratio AssessmentSignal-to-noise Ratio EstimationTurbulence CharacterizationTurbulence DetectionTurbulence MeasurementTurbulent Flow Measurement
An advanced version of the aerofoil probe turbulence gauge is described. The new device uses an axisymmetric lifting body sensor of 2.4 mm diameter coupled directly to a miniature preamplifier unit (6.35 mm diam). The output voltage varies linearly with the v or w component of turbulence. The device has uniform sensitivity and good signal-to-noise ratio over a nominal frequency range of 10–10 000 Hz. A discussion of calibration procedures and application considerations is included.