One-Step Detection of Pathogens and Viruses: Combining Magnetic Relaxation Switching and Magnetic Separation

195

Citations

26

References

2015

Year

Abstract

We report a sensing methodology that combines magnetic separation (MS) and magnetic relaxation switching (MS-MRS) for one-step detection of bacteria and viruses with high sensitivity and reproducibility. We first employ a magnetic field of 0.01 T to separate the magnetic beads of large size (250 nm in diameter) from those of small size (30 nm in diameter) and use the transverse relaxation time (T2) of the water molecules around the 30 nm magnetic beads (MB30) as the signal readout of the immunoassay. An MS-MRS sensor integrates target enrichment, extraction, and detection into one step, and the entire immunoassay can be completed within 30 min. Compared with a traditional MRS sensor, an MS-MRS sensor shows enhanced sensitivity, better reproducibility, and convenient operation, thus providing a promising platform for point-of-care testing.