NMR in Biomedicine · 2007 · 39 citations · 13 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMedical ImagingMedical Imaging InstrumentationMagnetic ImagingMagnetic Resonance Imaging Acquisition
The advantages of array coil imaging in human whole-body MR systems are well known and have gained relevance in many applications and MR techniques. In the field of small-animal studies, this concept has become increasingly important. In this work, three different phased-array coils for performing MRI on mice are presented. For imaging at 300 MHz, a four-channel receive-only phased-array coil is introduced. One two-channel and a four-channel transmit/receive setup operating at 750 MHz show the feasibility of array coil imaging at 17.6 T. All of these coils show excellent performance and deliver high-quality MR images of mice.
13
SENSE: Sensitivity encoding for fast MRI
Klaas P. Pruessmann, Markus Weiger, Markus B. Scheidegger et al. · Magnetic Resonance in Medicine · 1999 · 6.1K citations
Generalized autocalibrating partially parallel acquisitions (GRAPPA)
Mark A. Griswold, Peter M. Jakob, Robin M. Heidemann et al. · Magnetic Resonance in Medicine · 2002 · 5.3K citations · Full text
P. B. Roemer, W. A. Edelstein, Cecil E. Hayes et al. · Magnetic Resonance in Medicine · 1990 · 2.3K citations
FLASH imaging: Rapid NMR imaging using low flip-angle pulses
Axel Haase, Jens Frahm, D. Matthaei et al. · Journal of Magnetic Resonance · 2011 · 1.2K citations