A newly developed a-Se mammography flat panel detector with high-sensitivity and low image artifact

Yoshihiro Okada, Kei-ichiro Sato, Takaaki Ito, Yuichi Hosoi, Toshiro Hayakawa

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2013 · 10 citations · 0 references

Concepts

Abstract

A novel amorphous selenium (a-Se) detector with the hexagonal pixel has been developed for full-field digital mammography. The pixel area of the detector was designed to be same as that of the 68 &mu;m square pixel detector, while the pixel pitch between neighboring pixels was set to be 73-75 &mu;m in six directions. By applying the hexagonal pixels, the sensitivity of the detector has improved by 18% compared with a conventional square pixel. A simulation showed that the hexagonal pixel provided a more uniform electric field in the a-Se layer than the square pixel, which has lead to higher sensitivity. The modulation transfer function of the detector was 92 % at 2 mm<sup>-1</sup> and 62 % at 5 mm<sup>-1</sup>. These values were as high as that of a conventional a-Se detector with 50 &mu;m square pixels. As a result, the detective quantum efficiency of this detector achieved 75 % with 5 mR and 72 % with 2.5mR at 2 mm<sup>-1</sup>. The exposure conditions were 28 kV W/Rh with a 2 mm aluminum filter. Therefore, the new detector can reduce the exposure dose while maintaining a high image quality.