Concepedia

Publication | Closed Access

Stress Analysis of the Interface Between Cervical Vertebrae End Plates and the Bryan, Prestige LP, and ProDisc-C Cervical Disc Prostheses

106

Citations

10

References

2009

Year

Abstract

Rigidity of the cores ("nuclei") in Prestige LP and ProDisc-C prostheses guarantee initial maintenance of disc height, but high contact stress takes place at the bone-end plate interface if they are improperly placed or undersized. Anchorage designs add an additional factor that may increase propensity of subsidence, indicated by the high contact stress occurring at the end plate flanges of Prestige LP, and at midline keel fixation on the end plate of ProDisc-C. Although Bryan disc differs in these 2 concerns, it also creates much larger displacement during motion with more variation in disc height that may theoretically increase the load sharing of facet and/or uncovertebral joints compared to more rigid artificial discs.