Influence of absorbing and scattering media on the propagation of ultrasound

M. van den Aarssen, W. A. Verhoef, J.M. Thijssen

The Journal of the Acoustical Society of America · 1989 · 17 citations · 0 references

Concepts

Abstract

The preservation of the sound field produced by a single element ultrasound transducer was investigated for the assessment of the applicability of diffraction correction strategies. A variety of purely absorbing and absorbing plus scattering phantoms was employed. The acoustic parameters (velocity of sound, frequency-dependent attenuation coefficient, and differential backscattering coefficient) were estimated for these phantoms and compared to the parameters of human liver and skin specimens. The focal beam properties were estimated with a small diamond reflector positioned beyond the investigated media. The changes of the beam characteristics in the phantoms could be explained completely by the overall attenuation, i.e., either by absorption exclusively or by absorption plus scattering. In liver tissue, the averaged beam broadening could be explained by the attenuation, but great and irregular distortions of the sound beam were observed. The distortions by skin, subcutaneous fat, and muscle layers were even more severe. An important conclusion from the present study is, therefore, that diffraction correction strategies should be considered as a first-order approximation when applied to in vivo diagnostic procedures.