Publication | Closed Access
Intensity of the Stimulating Current May Not Be a Reliable Indicator of Intraneural Needle Placement
107
Citations
8
References
2008
Year
Specific response to nerve stimulation with currents <0.2 mA occurred only when the needle tip was positioned intraneurally. However, motor response could be absent with intraneural needle placement at a current intensity of up to 1.7 mA.