A switched-capacitor charge-balancing analog-to-digital converter and its application to capacitance measurement

Hiroki Matsumoto, Shimizu Hiromi, Kenzo Watanabe

IEEE Transactions on Instrumentation and Measurement · 1987 · 84 citations · 0 references

DOIFull text

Open access

Concepts

TL;DR

The authors develop a charge‑balancing analog‑to‑digital converter and demonstrate its applicability through selected applications. The converter uses a switched‑capacitor integrator, comparator, and digital logic driven by a two‑phase clock that accumulates signal charge while extracting reference charge, with the ratio of accumulated to extracted frequencies yielding the digital output. The integrated prototype achieves over 14‑bit accuracy, is insensitive to op‑amp offset, finite open‑loop gain, and parasitic capacitance, and occupies a very small chip area with few devices.

Abstract

An analog-to-digital converter is developed based on the charge-balancing principle. It consists of a switched-capacitor integrator, comparator, and digital logic circuit. Driven by the two phase clock, the integrator accumulates consecutively the incremental signal charge while extracting the quantized reference charge from the accumulated signal charge each time its output becomes positive, to keep their charge balance. The ratio between the accumulated and extracted frequencies for a given period of time then provides the digital representation of an input analog signal. A conversion accuracy higher than 14 bits can be expected from its integrated realization because the offset voltage and the finite open-loop gain of an op-amp and the parasitic capacitance have no effect upon the conversion process. It also features a small device-count integrate onto a very small chip area. Some applications are also presented to demonstrate its validity.