Neurosurgery · 1993 · 50 citations · 3 references
MedicineNeurological SurgeryNeurosurgeryBioinstrumentationResuscitationTraumatic Brain Injury PreventionHemodynamicsIntracranial HypertensionIntracranial PressureIntracranial Pressure ManagementIntracranial Pressure MonitoringTraumatic Brain InjuryCerebrospinal Fluid Pressure MeasurementContinuous Intracranial Pressure MonitoringIncreased Intracranial PressureIntracranial Pressure DynamicsIntracranial Pressure Measurement AccuracyIntracranial Pressure Monitoring DevicesIntracranial Pressure PhysiologyNoninvasive Intracranial Pressure Monitoring
The camino ventricular bolt system has been used to monitor intracranial pressure in patients after severe head injury. The correlation between the ventricular pressure measured with the Camino device and an external transducer showed that the Camino accurately measured intracranial pressure over a wide range, but that it read an average of 1.15 mm Hg higher than that obtained by the external transducer. The technique has the advantage over a remote transducer because it is sited within the ventricle. This may be of value in wave-form analysis.
3
A clinical evaluation of the Camino subdural screw and ventricular monitoring kits
I. R. Chambers, A. D. Mendelow, E J Sinar et al. · Neurosurgery · 1990 · 56 citations
A Clinical Evaluation of the Camino Subdural Screw and Ventricular Monitoring Kits
I. R. Chambers, A. David Mendelow, E J Sinar et al. · Neurosurgery · 1990 · 52 citations