Validation of a technique for estimating organ doses for kilovoltage cone-beam CT of the prostate using the PCXMC 2.0 patient dose calculator

Tim J Wood, C S Moore, John Saunderson, Andrew Beavis

Journal of Radiological Protection · 2015 · 18 citations · 29 references

Concepts

Abstract

The use of cone beam CT in common radiotherapy treatments is increasing with the growth of image guided radiotherapy. Whilst the benefits that this technology offers are clear, such as improved patient positioning prior to treatment, it is always important to consider the implications of such intensive imaging regimes on the patient, especially when considering the fundamental radiation protection requirements for justification and optimisation.The purpose of this study was to develop a technique that uses readily available dose calculation software (PCXMC 2.0) to estimate the organ and effective doses that result from these types of examination in prostate treatments on the Varian OBI system. It has been shown that by separating these types of examinations into 28 different projections, with a range of x-ray beam qualities, it is possible to reproduce the complex geometry that is used on these imaging systems in PCXMC i.e. asymmetric radiation field with a half bowtie filter rotating 360° around the patient.This new technique has been validated with thermo-luminescent dosimeter measurements in the Rando anthropomorphic phantom, and has been shown to give excellent agreement with this established method (R(2) = 0.995). This technique will prove to be valuable to radiotherapy departments that are looking to optimise their CBCT imaging protocols as it allows a rapid evaluation of the impact of any changes on patient dose. It also serves to further highlight the levels of dose that these types of patient are subject to when having daily CBCT scans as part of the treatment, which further reinforces the need for optimisation of both patient dose and image quality on these systems.

References

29