Journal of Clinical Ultrasound · 1982 · 64 citations · 15 references
MedicineCardiologyInstrumentationUltrasonographyUltrasoundDiagnostic UltrasonographyDoppler UltrasonographyDoppler UltrasoundTransducer TechnologyUltrasonicsUltrasound InstrumentationAnnular Array TransducersBlood Flow Doppler AssessmentClinical Laboratory EvaluationDoppler Blood Flow AssessmentDoppler Ultrasound InstrumentationSample Volume Measurement
Two moving string test targets have been developed and used to characterize ultrasound Doppler instrumentation. The characteristics investigated were sample volume size and location in space, amplitude sensitivity to Doppler shift, and resolution of Doppler direction. The tests were performed on a combined echo/Doppler instrument and on an annular array system. The procedures can be carried out routinely in the clinical laboratory to ensure that the instrument is working properly or as an aid for correct interpretation of acquired data.
15
Cardiovascular applications of ultrasound
P.N.T. Wells · Ultrasound in Medicine & Biology · 1975 · 309 citations
Medicine, Cardiology, Radiology +12
Ultrasonic Duplex Echo-Doppler Scanner
Frank E. Barber, Donald W. Baker, Arthur W. C. Nation et al. · IEEE Transactions on Biomedical Engineering · 1974 · 223 citations
Medicine, Cardiology, Radiology +17
The Dependence of Ultrasound Doppler Bandwidth on Beam Geometry
V.L. Newhouse, E.S. Furgason, Greg Johnson et al. · IEEE Transactions on Sonics and Ultrasonics · 1980 · 147 citations