Accurate dosimetry in scanning transmission X-ray microscopes<i>via</i>the cross-linking threshold dose of poly(methyl methacrylate)
28
Citations
51
References
2012
Year
X-ray SpectroscopyEngineeringMicroscopyBiomedical EngineeringAccurate DosimetryX-ray FluorescenceX-ray ImagingRadiation TestingX-ray TechnologyX-ray PolarizationInstrumentationRadiation ImagingRadiation OncologyNuclear MedicinePolymer ChemistryRadiologyHealth SciencesMaterials ScienceTransmission X-ray MicroscopesSoft X-raysSoft X-ray StxmsRadiographic ImagingDosimetryCross-linking Threshold DosePolymer ScienceX-ray DiffractionRadiation DoseX-ray Optic
The sensitivity of various polymers to radiation damage by soft X-rays has been measured previously with scanning transmission X-ray microscopes. However, the critical dose values reported by different groups for the same material differ by more than 100%. Possible sources of this variability are investigated here for poly(methyl methacrylate) (PMMA) using controlled exposure to monochromatic soft X-rays at 300 eV. Radiation sensitivity, judged by several different criteria, was evaluated as a function of dose rate, pre-exposure thermal treatments and X-ray polarization. Both the measured critical dose and the dose required to initiate negative mode (cross-linking) were observed to depend only on dose, not the other factors explored. A method of determining detector efficiency from the dose required to initiate negative mode in PMMA is outlined. This method was applied to many of the soft X-ray STXMs presently operating to derive the efficiencies of their transmitted X-ray detectors in the C 1s absorption-edge region.