Artificial Organs · 1994 · 11 citations · 3 references
ComputingVisualizationFlow VisualizationFluid DynamicsHydraulicsFlow MeasurementLaminar FlowTurbulence VisualizationUnsteady Flow VisualizationVortex DynamicsVortex VisualizationExperimental Visualization TechniquesExperimental Vortex VisualizationImpeller HydrodynamicsIntravascular HemolysisThrombus RemovalThrombus ResolutionVortex FlowVortex Formation
A compact centrifugal pump, NCVC-1, has an open-type impeller with 6 curved vanes, and it is characterized by no shaft and no seal. A tunnel is placed in the center of the impeller-rotor assembly to irrigate the back space behind the rotor. To evaluate the flow, we performed 3 visualization methods: tracer, oil film, and injection streak line method. The flow, observed by the tracer method in NCVC-1, indicated little turbulence along vanes. A volute chamber proved effective to reduce vortex formation in the outlet. Oil film pattern revealed no flow separation on vanes at 5 L/min. Washout flow behind the rotor is essential to prevent thrombus formation and was shown as inward spiral flow without any stagnation. These data suggested that a combination of visualization techniques was useful to analyze various flow conditions, and the NCVC-1 has excellent flow characteristics with little turbulence and little flow stagnation, which must be beneficial to low hemolysis and high antithrombogenicity.
3
Investigation of the flow in a centrifugal blood pump.
K. Affeld, K. Schichl, Ajit Yoganathan · PubMed · 1987 · 10 citations
Engineering, Physics, Physiology +12