Catheterization and Cardiovascular Interventions · 2001 · 19 citations · 15 references
MedicineCardiologySurgeryCardiothoracic SurgeryInterventional CardiologyVascular SurgeryBalloon AngioplastyCardiac CatheterizationCoronary AngioplastyHemodynamicsPercutaneous Coronary InterventionStent PlacementBalloon-expandable StentsBlood FlowCoronary Stent AssessmentCoronary Stent DeploymentPulsatile FlowPulsatile Flow ExperimentsStent Patency AssessmentVascular Stenting
We tested whether fractional flow reserve (FFR) discriminates between suboptimally and optimally deployed stents. Latex tubes (diameter solidus in circle = 4 mm) with diameter stenosis 40% (n = 3), 50% (n = 3) and 60% (n = 3) were tested in a pulsatile flow system, using water. Measurements were done at baseline (n = 9; FFR/QCA) and after suboptimal (SOD; 3-mm balloon at 8 atm) and optimal (OD; 4 mm balloon at 16 atm) deployment of a 35-mm stent (n = 6; FFR/QCA/IVUS). Varying Q from 150 to 50 ml/min increased FFR by 2-7%. Conversely, at 100 ml/min, FFR increased by only 0.8% from SOD to OD (P < 0.05). Extrapolating data to blood flow, the gain in FFR from SOD to OD is less than 5% for Q = 100 ml/min, while FFR may increase by 15-20% by changes in blood flow from 50 to 150 ml/min. We conclude that IVUS and QCA are more appropriate for the assessment of optimal stent deployment.
15
Patrick W. Serruys, Peter de Jaegere, Ferdinand Kiemeneij et al. · New England Journal of Medicine · 1994 · 4.5K citations · Full text
N. H. J. Pijls, J A van Son, R L Kirkeeide et al. · Circulation · 1993 · 1.2K citations
Medicine, Cardiology, Physiology +17
Victor L. Streeter, J. Kestin · Journal of Applied Mechanics · 1961 · 558 citations · Full text