2007 · 54 citations · 3 references
EngineeringBioengineeringElectrical EngineeringAnalog ElectronicsBiosensorsNanobiotechnologyAnalog InstrumentationBiological SensorsBiomedical SensorsBiopotential MeasurementBiosensor DesignNanosensorsBiosensor Signal AmplificationCurrent Sensing CircuitsFeedbackLow-noise Analog InterfacesNanobiosensorsTransimpedance Amplifier DesignTransimpedance Amplifiers
Active resistors up to 300GOmega operating down to the fA range are implemented in the feedback path of an integrator-differentiator transimpedance amplifier for high-sensitivity current measurements. The system has 4fA/radicHz noise up to 100kHz, a 2MHz bandwidth, and ensures unlimited measuring time even with nA DC input currents
3
Mark Akeson, Daniel Branton, John J. Kasianowicz et al. · Biophysical Journal · 1999 · 951 citations · Full text