Detection of aortic insufficiency by pulse doppler echocardiography

Jonathan M. Ward, Donald W. Baker, Simeon A. Rubenstein, Steve Johnson

Journal of Clinical Ultrasound · 1977 · 53 citations · 13 references

Concepts

Abstract

Sixty-five patients were studied for the presence of aortic insufficiency by auscultation, conventional M mode echocardiography (echo), range-gated pulse Doppler flowmeter (Doppler), and supravalvular aortic angiography. Aortic insufficiency was demonstrated angiographically in 49 patients, of which eight (16%) were missed by auscultation. Echo missed 20 (41%), and Doppler missed seven (14%). Only two (4%) were missed by auscultation and Doppler combined. All methods were highly specific, but the combination of auscultation and Doppler was the most sensitive noninvasive method for detecting aortic insufficiency.

References

13